胡海曼,雷雨蒙,余靖,王琦,雷炳松,叶华容,张舸.奇异值分解滤波联合Frangi滤波对超微血管成像图像质量的影响[J].中国医学影像技术,2024,40(3):430~435 |
奇异值分解滤波联合Frangi滤波对超微血管成像图像质量的影响 |
Impact of singular value decomposition filter combined with Frangi filter on image quality of superb microvascular imaging |
投稿时间:2023-11-21 修订日期:2023-12-28 |
DOI:10.13929/j.issn.1003-3289.2024.03.023 |
中文关键词: 血管瘤 超声检查 图像质量 奇异值分解 Frangi滤波器 |
英文关键词:hemangioma ultrasonography image quality singular value decomposition Frangi filter |
基金项目:湖北省自然科学基金青年项目(2023AFB289)。 |
作者 | 单位 | E-mail | 胡海曼 | 武汉亚洲心脏病医院心血管内科, 湖北 武汉 430022 华润武钢总医院超声医学科, 湖北 武汉 430080 湖北工业大学电气与电子工程学院, 湖北 武汉 430068 | | 雷雨蒙 | 华润武钢总医院超声医学科, 湖北 武汉 430080 | | 余靖 | 华润武钢总医院超声医学科, 湖北 武汉 430080 | | 王琦 | 华润武钢总医院超声医学科, 湖北 武汉 430080 | | 雷炳松 | 华润武钢总医院超声医学科, 湖北 武汉 430080 | | 叶华容 | 华润武钢总医院超声医学科, 湖北 武汉 430080 | | 张舸 | 武汉亚洲心脏病医院心血管内科, 湖北 武汉 430022 华润武钢总医院超声医学科, 湖北 武汉 430080 | 451170880@qq.com |
|
摘要点击次数: 183 |
全文下载次数: 157 |
中文摘要: |
目的 观察奇异值分解(SVD)滤波联合Frangi滤波对超微血管成像(SMI)图像质量的影响。方法 分别对1具仿体及1例疑诊肝血管瘤患者行CEUS,分别以SVD滤波(A组)、Frangi滤波(B组)、Frangi+SVD滤波(C组)及SVD+Frangi滤波(D组)处理图像,之后比较不同图像的质量。结果 4组仿体及人体肝脏SMI对比组织比(CTR)、对比度噪声比(CNR)及信噪比(SNR)均高于原始CEUS图像。相比原始图像,D组图像质量及分辨率提升最为显著,仿体SMI的 CTR、CNR及SNR分别提升58.04、3.39及48.04 dB,人体肝脏SMI相应参数则分别提升61.85、16.80及49.67 dB,而分辨率分别为原始CEUS图像的1.42倍及1.98倍。结论 SVD滤波联合Frangi滤波可有效提高SMI图像质量。 |
英文摘要: |
Objective To observe the impact of singular value decomposition (SVD) filter combined with Frangi filter on image quality of superb microvascular imaging (SMI). Methods Contrast-enhanced ultrasound (CEUS) was performed in a phantom and a patient with suspected liver hemangioma. Images in group A were processed with SVD filter, while in group B, C and D with Frangi filter, Frangi+SVD filter and SVD+Frangi filter, respectively, then image quality of SMI were compared. Results Contrast-to-tissue ratio (CTR), contrast-to-noise ratio (CNR) and signal-to-noise ratio (SNR) of A,B, C and D group were all higher than those of original CEUS images. Compared with the original CEUS images, the image quality indexes and image resolution of group D were most significantly improved, with CTR, CNR and SNR increased by 58.04, 3.39 and 48.04 dB of the phantom, as well as increased by 61.85, 16.80 and 49.67 dB of the patient, respectively. The resolution of the phantom's and the patient's SMI images was 1.42 times and 1.98 times of the original CEUS images, respectively. Conclusion SVD filter combined with Frangi filter could effectively improve image quality of SMI. |
查看全文 查看/发表评论 下载PDF阅读器 |
|
|
|