江建芹,蔡荣芳,崔磊,尹剑兵,许一鸣,马航,何海艳.肺癌体素内不相干运动参数值测量的短期可重复性及影响因素[J].中国医学影像技术,2018,34(4):543~547
肺癌体素内不相干运动参数值测量的短期可重复性及影响因素
Short-term reproducibility and impact factors of intravoxel incoherent motion parameters for lung cancer
投稿时间:2017-07-27  修订日期:2018-01-25
DOI:10.13929/j.1003-3289.201707142
中文关键词:  肺肿瘤  扩散磁共振成像  体素内不相干运动  可重复性,结果
英文关键词:Lung neoplasms  Diffusion magnetic resonance imaging  Intravoxel incoherent motion  Reproducibility of results
基金项目:江苏省2017年高层次卫生人才"六个一工程"拔尖人才项目(LGY2017037)、南通市科技局医技综合研究重点专项课题、盐城市医学科技发展计划项目(YK2017021)。
作者单位E-mail
江建芹 盐城市第一人民医院影像科, 江苏 盐城 224001  
蔡荣芳 南通大学第二附属医院影像科, 江苏 南通 226001  
崔磊 南通大学第二附属医院影像科, 江苏 南通 226001 cuigeleili@126.com 
尹剑兵 南通大学第二附属医院影像科, 江苏 南通 226001  
许一鸣 南通大学第二附属医院胸外科, 江苏 南通 226001  
马航 南通大学第二附属医院呼吸科, 江苏 南通 226001  
何海艳 南通大学第二附属医院呼吸科, 江苏 南通 226001  
摘要点击次数: 1784
全文下载次数: 1032
中文摘要:
      目的 评估肺癌短期重复扫描体素内不相干运动(IVIM)参数值测量的短期可重复性及影响因素。方法 对38例肺癌患者进行2次自由呼吸IVIM成像(间隔0.5~1.0 h)。由2名测量者基于IVIM图像独立测量慢扩散参数(D)、快扩散系数(D*)及灌注分数(f)均值。采用配对样本t检验或Wilcoxon检验比较重复测量和重复扫描的误差。以组内相关系数(ICC)评价各参数值同一测量者和不同测量者的一致性和可重复性。采用变异系数(WCV)比较肺癌类型(周围型、中央型)、病变大小(≥ 2 cm、<2 cm)及位置(上、中、下肺野)对各参数值可重复性的影响。结果 所有参数2次测量的系统误差差异均无统计学意义(P均>0.05)。D*和f值的WCVs(19.88%~36.83%)明显高于D值(2.97%~4.62%)。D值在中央型肺癌及最大径< 2 cm病灶的WCVs较高,并自上肺野到下肺野WCV依次增加;f在中央型肺癌和最大径≥ 2 cm的WCVs较大;D*受肺癌类型、大小和位置的影响显著。结论 在肺癌的重复扫描中,D值可重复性好,而D*、f值的可重复性较差;肺癌类型、大小和位置是影响IVIM参数值的因素。
英文摘要:
      Objective To prospectively evaluate the short-term test-retest reproducibility of intravoxel incoherent motion (IVIM) parameters for lung cancer, and to investigate its impact factors. Methods Thirty-eight lung cancer patients underwent twice free-breathing IVIM scanning (0.5-1.0 h interval). Two radiologists independently analyzed IVIM images to obtain true diffusion coefficient (D), pseudo-diffusion coefficient (D*) and perfusion fraction (f). All the parameters were compared between repeated measurements and scans. Intra-and inter-observer, test-retest reproducibility was assessed with interclass correlation coefficients (ICCs). The impact of lung cancer type (peripheral and central), size (the largest diameter ≥ 2 cm and <2 cm) and location (upper, mid, and lower lung field) on short-term reproducibility were compared by using coefficient of variations (WCVs). Results There was no significant difference between repeated measurements for all parameters (all P>0.05). All the parameters showed good intra-and inter-observer agreement. WCVs of D* and f (19.88%-36.83%) were higher than those of D (2.97%-4.62%). WCVs of D were higher for central lung cancer and the largest diameter <2 cm, and increasing from apical to lower zone. WCV of f was larger in the central lung cancer and the largest diameter ≥ 2 cm, while D* was greatly influenced by type, size and location of lung cancer. Conclusion The test-retest reproducibility of D is good, while of D* and f are poor in lung cancer. The type, size and location of lung cancers are the impact factors of IVIM parameters.