孙岩,沈均康,郝跃峰,秦卫,徐吉雄,胡丹,刘可夫.磁共振小视野表面线圈在距骨软骨损伤成像中的应用[J].中国医学影像技术,2017,33(3):454~457
磁共振小视野表面线圈在距骨软骨损伤成像中的应用
Application of MR small FOV surface coil in talar osteochondral lesions imaging
投稿时间:2016-08-05  修订日期:2016-12-16
DOI:10.13929/j.1003-3289.201608027
中文关键词:  磁共振成像  射频线圈  软骨损伤  距骨
英文关键词:Magnetic resonance imaging  Radiofrequency coil  Osteochondral lesion  Talus
基金项目:苏州市临床重点病种诊疗技术专项项目(LCZX201310)。
作者单位E-mail
孙岩 南京医科大学附属苏州医院放射科, 江苏 苏州 215008  
沈均康 苏州大学附属第二医院放射科, 江苏 苏州 215004  
郝跃峰 南京医科大学附属苏州医院骨科, 江苏 苏州 215008  
秦卫 南京医科大学附属苏州医院骨科, 江苏 苏州 215008  
徐吉雄 无锡人民医院放射科, 江苏 无锡 214023 70854598@qq.com 
胡丹 南京医科大学附属苏州医院骨科, 江苏 苏州 215008  
刘可夫 南京医科大学附属苏州医院放射科, 江苏 苏州 215008  
摘要点击次数: 2024
全文下载次数: 1006
中文摘要:
      目的 比较靴形线圈的质子密度加权成像(2D-FSE-PD)、三维可变翻转角快速自旋回波序列(3D-SPACE)成像和小FOV表面线圈2D-FSE-PD序列成像对距骨软骨损伤成像的准确率。方法 对 43例患者(45个踝关节)采用靴形线圈行2D-FSE-PD和3D-SPACE序列扫描,然后更换为小FOV表面线圈行2D-FSE-PD成像。以关节镜检查结果为金标准,比较3种扫描方法显示距骨软骨损伤的符合率。结果 小FOV表面线圈2D-FSE-PD序列的符合率为86.67%(39/45),靴形线圈2D-FSE-PD序列的符合率为60.00%(27/45),靴形线圈3D-SPACE序列的符合率为68.89%(31/45)。小FOV表面线圈2D-FSE-PD序列的符合率高于靴形线圈3D-SPACE序列(χ2=4.114,P=0.002)和2D-FSE-PD序列(χ2=8.182,P<0.001),而靴形线圈3D-SPACE序列的符合率与2D-FSE-PD序列的差异无统计学意义(χ2=0.776,P=0.125)。结论 与靴形线圈的2D-FSE-PD与3D-SPACE序列相比,小FOV表面线圈2D-FSE-PD成像能更好地显示距骨软骨损伤。
英文摘要:
      Objective To compare the accuracy rate of talar osteochondral lesions between the small field of view (FOV) surface coil MR imaging and 2D FSE proton density weighted imaging (2D-FSE-PD), 3D-SPACE sequences MR imaging with boot-shaped coil. Methods Totally 43 patients (45 ankles) underwent 3D-SPACE sequence after 2D-FSE-PD sequence scanning using the boot-shaped coil, then the PDWI sequence was scanning after the boot-shaped coil was replaced by the small FOV surface coil at 1.5T MRI. The results of the arthroscopic examinations was regarded as the standard for assessing the accuracy rate of the three methods. Results The accuracy rate of 2D-FSE-PD sequence with small FOV surface coil was 86.67% (39/45), the accuracy rate of 2D-FSE-PD sequence with boot coil was 60.00% (27/45), the accuracy rate of 3D-SPACE sequence with boot coil was 68.89% (31/45). The accuracy rate of small FOV surface coil was higher than those of boot coil used 3D-SPACE sequence (χ2=4.114, P=0.002) and boot coil used 2D-FSE-PD sequence (χ2=8.182, P<0.001). There was not significant difference between 3D-SPACE sequence using boot coil and 2D-FSE-PD sequence using boot coil (χ2=0.776, P=0.125). Conclusion Compared with the 2D-FSE-PD and 3D-SPACE sequences with boot-shaped coil, the small surface coil scanning with routine sequence can show talar osteochondral lesions better.