康泰山,林良杰,吴健,林建忠,徐丽娜.基于mDixon新型MR血管成像序列用于颈部动脉成像的可行性[J].中国医学影像技术,2023,39(4):592~596
基于mDixon新型MR血管成像序列用于颈部动脉成像的可行性
Feasibility of a novel MR angiography sequence based on mDixon for cervical artery imaging
投稿时间:2022-11-22  修订日期:2023-01-27
DOI:10.13929/j.issn.1003-3289.2023.04.024
中文关键词:  颈动脉疾病  磁共振血管成像  水脂分离技术  压缩感知
英文关键词:carotid artery diseases  magnetic resonance angiography  mDixon technology  compressed sensing
基金项目:中国博士后科学基金会面上项目(2020M671947)。
作者单位E-mail
康泰山 厦门大学附属中山医院磁共振科, 福建 厦门 361004
厦门市放射质量控制中心, 福建 厦门 361004 
 
林良杰 飞利浦(中国)投资有限公司北京分公司, 北京 100016  
吴健 厦门大学电子科学与技术学院电子科学系, 福建 厦门 361005  
林建忠 厦门大学附属中山医院磁共振科, 福建 厦门 361004
厦门市放射质量控制中心, 福建 厦门 361004 
 
徐丽娜 厦门大学电子科学与技术学院电子科学系, 福建 厦门 361005 lnx@xmu.edu.cn 
摘要点击次数: 1613
全文下载次数: 823
中文摘要:
      目的 基于将放射状k空间填充及压缩感知技术融合于mDixon技术开发一种新型MR血管成像(MRA)序列(3D mDixon-MRA),观察其用于颈部动脉成像的可行性。方法 前瞻性收集45例疑诊颈部动脉疾病患者,以3D时间飞跃法采集颈部动脉MRA(3D TOF-MRA)及3D mDixon-MRA,比较颈总动脉根部和中远段、椎动脉根部和中远段及颈内外动脉根部在2种图像中的信噪比(SNR)和对比度噪声比(CNR),以及2种图像质量主观评分的差异。结果 3D mDixon-MRA所示各动脉SNR均显著高于3D TOF-MRA(P均<0.01)。颈总动脉根部CNR在2种图像中的差异无统计学意义(P=0.13),其余动脉在3D mDixon-MRA中的CNR均显著高于3D TOF-MRA(P均<0.01)。颈总动脉根部、椎动脉根部及椎动脉中远段在3D mDixon-MRA中的主观评分均高于3D TOF-MRA(P均<0.05),其余动脉在2种图像中的主观评分差异均无统计学意义(P均>0.05)。结论 3D mDixon-MRA序列可用于颈部动脉成像,有利于提高图像质量、缩短扫描时间。
英文摘要:
      Objective To develop a novel MR angiography sequence (3D mDixon-MRA) based on integrating k-space radial sampling and compressed sensing technology into mDixon technique, and to explore its feasibility for application in cervical artery imaging. Methods Forty-five patients with suspected cervical artery diseases were prospectively enrolled, and 3D time-of-flight cervical artery MRA (3D TOF-MRA) and 3D mDixon-MRA were performed. Signal-to-noise ratio (SNR) and contrast-to-nosie ratio (CNR) of root and mid-distal of the common carotid artery, root and mid-distal of the vertebral artery, root of internal and external carotid arteries, as well as the subjective scores of imaging quality were compared between 2 kinds of images. Results SNR of each artery on 3D mDixon-MRA was significantly higher than that on 3D TOF-MRA (all P<0.01). No significant difference of CNR of common carotid artery root was found between 3D mDixon-MRA and 3D TOF-MRA (P=0.13), while CNR of the other arteries were significantly higher on 3D mDixon-MRA than those on 3D TOF-MRA (all P<0.01). The subjective scores of common carotid artery root, the root and mid-distal of vertebral artery on 3D mDixon-MRA were significantly higher than those on 3D TOF-MRA (all P<0.05), while no significant difference of subjective scores of the rest arteries was found between 2 kinds of images (all P>0.05). Conclusion 3D mDixon-MRA sequence could be used for cervical artery imaging, which could improve imaging quality and shorten scanning time.