于安宏,杨雅朋,马延贺,高洪岩,张洪.超低剂量CT能谱纯化技术联合高级建模迭代重建对显示肺磨玻璃结节及测量其体积的影响:体模研究[J].中国医学影像技术,2022,38(5):764~769
超低剂量CT能谱纯化技术联合高级建模迭代重建对显示肺磨玻璃结节及测量其体积的影响:体模研究
Impact of tin filtration combining advanced modeled iterative reconstruction on imaging quality of ultra-low dose CT and measuring volume of lung ground glass nodules: A phantom study
投稿时间:2021-08-09  修订日期:2021-12-27
DOI:10.13929/j.issn.1003-3289.2022.05.030
中文关键词:  肺肿瘤  体模,显像术  胸部  体层摄影术,X线计算机  迭代重建  能谱纯化
英文关键词:lung neoplasms  phantoms, imaging  thorax  tomography, X-ray computed  iterative reconstruction  spectral filtration
基金项目:天津市科技计划(18ZXZNSY00400)。
作者单位
于安宏 天津市胸科医院影像科, 天津 300222 
杨雅朋 天津市胸科医院影像科, 天津 300222 
马延贺 天津市胸科医院影像科, 天津 300222 
高洪岩 天津市胸科医院影像科, 天津 300222 
张洪 天津市胸科医院影像科, 天津 300222 
摘要点击次数: 2234
全文下载次数: 469
中文摘要:
      目的 观察超低剂量CT能谱纯化技术联合高级建模迭代重建(ADMIRE)对显示肺磨玻璃结节(GGN)及测量其体积的影响。方法 将20枚模拟GGN随机置于模拟成年男性的胸部体模肺中,以常规低剂量(A方案)及管电压Sn 100 kV能谱纯化技术(B方案)行CT扫描,并以滤波反投影法(FBP)及ADMIRE(强度1~5)进行重建;通过观察模拟GGN的CT值标准差(SD)、体积、体积相对误差(RVE)、主观图像质量评分及辐射剂量,评估B方案的价值。结果 B方案各重建方式模拟肺GGN CT值SD均高于A方案(P均<0.05);相同方案不同重建方式所得CT值SD差异均有统计学意义(P均<0.05)。以相同方案不同重建方式模拟肺GGN CT图像的主观质量评分由高到低依次为ADMIRE-4、ADMIRE-3、ADMIRE-5、ADMIRE-2、ADMIRE-1及FBP;相同重建方式下,B方案所获CT图像主观质量评分均低于A方案(P均<0.05)。相同方案下,不同重建方式模拟GGN RVE差异均有统计学意义(P均<0.001),且基本随FBP及ADMIRE强度增高而呈递减趋势。以FBP、ADMIRE-1、ADMIRE-2及ADMIRE-3重建图像时,B方案模拟GGN 的RVE均大于A方案(P均<0.05)。A、B方案的有效辐射剂量分别为0.51 mSv及0.12 mSv。结论 采用管电压Sn 100 kV超低剂量CT能谱纯化技术结合ADMIRE-4检测体模GGN,可在降低辐射剂量的同时维持图像质量及测量其体积的精确度。
英文摘要:
      Objective To observe the impact of tin filtration combining advanced modeled iterative reconstruction (ADMIRE) on image quality of ultra-low dose CT and measuring volume of lung ground glass nodule (GGN).Methods Twenty simulated GGN were randomly placed in the lungs of a chest phantom imitating adult male. CT scanning was performed with conventional low-dose (scheme A) and tin filtration with tube voltage Sn 100 kV (scheme B), both reconstructed using filtered back projection (FBP) and ADMIRE (strength level from 1 to 5). The standard deviation (SD) of CT value, the volume, relative volume error (RVE), the subjective image quality score and radiation dose of simulated lung GGN under different scanning schemes and reconstruction modes were observed, and the value of scheme B was explored.Results SD of CT value of simulated GGN with each reconstruction mode under scheme B was higher than those of scheme A (all P<0.05). Under the same scheme, there were significant differences of SD of CT values obtained with different reconstruction modes (all P<0.05). Under the same scheme, the subjective quality scores of simulated lung GGN on CT images obtained with different reconstruction modes from high to low were ADMIRE-4, ADMIRE-3, ADMIRE-5, ADMIRE-2, ADMIRE-1 and FBP, while with the same reconstruction mode, the subjective quality scores of CT images under scheme B were lower than those under scheme A (all P<0.05). Under the same scheme, there were significant differences of simulated GGN RVE obtained with different reconstruction modes (all P<0.001), which basically decreased with FBP and ADMIRE strength increased. When reconstruction with FBP, ADMIRE-1, ADMIRE-2 and ADMIRE-3, the simulated GGN RVE of scheme B was higher than those of scheme A (all P<0.05). The effective radiation dose of scheme A and B was 0.51 mSv and 0.12 mSv, respectively.Conclusion Tin filtration with tube voltage Sn 100 kV ultra-low dose CT combined with ADMIRE-4 could effectively reduce the radiation dose and remain CT image quality as well as the accuracy of detecting measuring volume of phantom GGN.