柴亚如,邢静静,高剑波,吕培杰,侯平,王素雅,刘杰.多模型迭代重建算法对腹部体模CT扫描图像质量和辐射剂量的影响[J].中国医学影像技术,2018,34(1):118~122
多模型迭代重建算法对腹部体模CT扫描图像质量和辐射剂量的影响
Influence of adaptive statistical iterative reconstruction-V on image quality and radiation dose of abdominal CT in phantom
投稿时间:2017-04-10  修订日期:2017-10-20
DOI:10.13929/j.1003-3289.201704047
中文关键词:  体层摄影术,X线计算机  迭代重建  辐射剂量  图像质量  体模
英文关键词:Tomography,X-ray computed  Iterative reconstruction  Radiation dosage  Image quality  Phantom
基金项目:国家自然科学基金面上项目(81671682)、郑州大学第一附属医院青年创业基金([2012]233号)。
作者单位E-mail
柴亚如 郑州大学第一附属医院放射科, 河南 郑州 450052  
邢静静 郑州大学第一附属医院放射科, 河南 郑州 450052  
高剑波 郑州大学第一附属医院放射科, 河南 郑州 450052 cjr.gaojianbo@vip.163.com 
吕培杰 郑州大学第一附属医院放射科, 河南 郑州 450052  
侯平 郑州大学第一附属医院放射科, 河南 郑州 450052  
王素雅 郑州大学第一附属医院放射科, 河南 郑州 450052  
刘杰 郑州大学第一附属医院放射科, 河南 郑州 450052  
摘要点击次数: 2706
全文下载次数: 1114
中文摘要:
      目的 分析前置全模型迭代重建算法(ASIR-V)在腹部体模CT扫描中图像质量的变化规律及辐射剂量的降低程度,探讨优化图像质量和辐射剂量的最佳前置ASIR-V。方法 采用腹部仿真体模和Revolution CT机。根据噪声指数(NI)设置(6、8、10、12、14)分为5组。每组设置0~100%(间隔10%)前置ASIR-V扫描和常规扫描(即不联合迭代),共获得55组图像。分析各NI组图像CT值、噪声、主观评分及辐射剂量随ASIR-V比例的变化规律。各NI组图像的主观评分比较采用秩和检验,CT值、噪声和辐射剂量比较采用单因素方差分析和配对t检验。结果 在0~40%前置ASIR-V水平,NI为6、8、10组图像主观评分基本稳定,NI为12、14组图像主观评分呈略升高趋势;50%~100%前置ASIR-V,各组主观评分均呈逐渐降低趋势;NI为6、8、10组,超过70%前置ASIR-V图像主观评分降至3分以下,NI为12、14组,超过60%前置ASIR-V主观评分降至3分以下;NI为6、8、10组,常规扫描图像主观评分与40%前置ASIR-V图像差异无统计学意义(P=0.626、0.915、0.514),NI为12、14组,常规扫描图像主观评分略低于40%前置ASIR-V图像(P=0.041、0.036);各NI组常规扫描图像主观评分均优于60%前置ASIR-V图像((P=0.021、0.012、0.015、0.014、0.007))。各NI组中不同部位CT值、噪声随前置ASIR-V比例的增高呈基本稳定状态(P均>0.05)。各NI组CTDIvol随前置ASIR-V比例的增高,呈逐渐下降趋势。40%、50%和60%前置ASIR-V比例条件下,CTDIvol较常规扫描组下降比率分别为49.82%、62.51%、71.63%。结论 前置ASIR-V可在保证图像质量的条件下明显降低辐射剂量;腹部前置ASIR-V比例推荐40%~60%。
英文摘要:
      Objective To evaluate the influence of preset adaptive statistical iterative reconstruction-V (ASIR-V) techniques on image quality and radiation dose reduction of abdominal CT in phantom, and to investigate the optimal ASIR-V level.Methods Abdominal anthropomorphic phantom was scanned using Revolution CT, when noise index (NI) were set as 6, 8, 10, 12 and 14, respectively. Then 0-100% ASIR-V and conventional scan was performed and 55 sets of images were obtained. CT value, noise, subjective score and radiation dose were recorded, and the optimal ASIR-V was obtained. Subjective scores of images in each group were compared using rank sum test, and CT value, noise and radiation dose were compared with one way ANOVA and paired t test.Results The image subjective score unchanged when NI was 6, 8 or 10, slightly increased when NI was 12 and 14 with 0-40% ASIR-V, and decreased above 50% ASIR-V at all NI. When NI was 6, 8 or 10, more than 70% ASIR-V image subjective score fell below 3 points. When NI was 12 or 14 group, more than 60% ASIR-V subjective score fell below 3 points. The image quality score of conventional scan had no difference with 40% ASIR-V when NI was 6, 8 or 10, respectively (P=0.626, 0.915, 0.514), and inferior to 40% ASIR-V when NI was 12 or 14 (P=0.041, 0.036), while in all NI group, image quality score of conventional scan was superior to 60% ASIR-V (P=0.021, 0.012, 0.015, 0.014, 0.007). CT values and image noises had no significant differences in different parts in all NI groups (all P>0.05). CT dose index volume (CTDIvol) continuing decreased with ASIR-V. Compared with that of conventional scan, at 40%, 50% and 60% ASIR-V, CTDIvol reduced by 49.82%, 62.51% and 71.63%, respectively.Conclusion Preset ASIR-V can reduce radiation dose obviously while maintaining the overall image quality, and 40%-60% ASIR-V can be recommended for abdominal CT in clinical application.