冉姗姗,綦维维,张淼,张卓璐,陈雷,洪楠.多参数设置对低剂量胸部CT扫描图像质量及辐射剂量的影响[J].中国医学影像技术,2018,34(1):113~117 |
多参数设置对低剂量胸部CT扫描图像质量及辐射剂量的影响 |
Impact of multiple parameters setting on image quality and radiation dose in low-dose chest CT scanning |
投稿时间:2016-04-10 修订日期:2017-11-16 |
DOI:10.13929/j.1003-3289.201704039 |
中文关键词: 胸部 体层摄影术,X线计算机 辐射剂量 图像质量 |
英文关键词:Chest Tomography,X-ray computed Radiation dosage Image quality |
基金项目:北京市科委早期肺癌筛查项目(D141100000214006)。 |
|
摘要点击次数: 3989 |
全文下载次数: 2291 |
中文摘要: |
目的 探讨当iCT管电流设置为17 mAs时,其他参数设置对低剂量胸部CT扫描图像质量及辐射剂量的影响。方法 选取低剂量胸部CT筛查的200名志愿者,设定管电流为17 mAs,根据管电压及自动曝光控制(DoseRight)设置分为4组:A组为120 kV,DoseRight扫描;B组为100 kV,DoseRight扫描;C组为120 kV,固定管电流扫描;D组为100 kV,固定管电流扫描。对所有扫描图像分别采用混合迭代重建(iDose4)和全模型迭代重建(IMR)进行肺窗及纵隔窗算法重建。比较4组的有效剂量及图像质量。结果 4组肺窗、纵隔窗的iDose4、IMR重建图像质量评分及有效剂量差异均有统计学意义;A、B组图像质量均较C组图像质量好(P均<0.05),与其余3组比较D组图像质量最差(P均<0.05);C组纵隔窗及D组肺窗、纵隔窗的iDose4重建图像质量评分较低;A、B、C、D组SNR值依次降低(P均<0.01);A组有效剂量最高(P均<0.01),D组有效剂量最低(P均<0.01),B组和C组间有效剂量差异无统计学意义(P=0.055)。结论 iCT管电流为17 mAs时,采用100 kV管电压、DoseRight扫描及IMR重建方式可获得较满意的图像质量和较低的辐射剂量。 |
英文摘要: |
Objective To investigate the impact of parameters setting on image quality and radiation dose of low-dose chest CT scanning when iCT tube current was set as 17 mAs.Methods Two hundred volunteers who accepted low-dose chest CT screening with tube current was set as 17 mAs were randomly divided into four groups according to the tube voltage and automatic exposure control (DoseRight) settings (group A:120 kV, DoseRight; group B:100 kV, DoseRight; group C:120 kV, fixed tube current; group D:100 kV, fixed tube current). Then hybrid iterative reconstruction (HIR:iDose4) and iterative model reconstruction (IMR) were performed on all scanned images of lung and mediastinal window algorithm, respectively. The effective dose and image quality of 4 groups were compared.Results There were significant differences of image quality score and effective dose among 4 groups. The image quality of group A and B were better than that of group C (P<0.05), and of group D was the worst compared with the other three groups (P<0.05). Image quality scores of the mediastinal window of group C and D and the lung window of group D at iDose4 reconstruction were lower. SNR value of group A, B, C, D decreased in order (P<0.01). Group A had the highest effective dose (P<0.01), while the effective dose of group B and group C had no significant difference (P=0.055), and group D had the lowest effective dose (P<0.01).Conclusion When tube current was set as 17mAs at iCT, selecting 100 kV tube voltage, DoseRight and IMR reconstruction arithmetic can obtain satisfactory image quality and low radiation dose. |
查看全文 查看/发表评论 下载PDF阅读器 |
|
|
|