刘宏,万泰虎,王永亮,赵卓,柳林,乔月,张孟超.能谱CT物质分离技术克服肾囊肿伪强化的临床评估[J].中国医学影像技术,2017,33(6):897~901
能谱CT物质分离技术克服肾囊肿伪强化的临床评估
Clinical assessment of spectrum CT with material decomposition technique in overcoming renal cyst pseudoenhancement
投稿时间:2016-10-16  修订日期:2017-03-03
DOI:10.13929/j.1003-3289.201610060
中文关键词:  体层摄影术,X线计算机  物质分离技术  肾囊肿  伪强化
英文关键词:Tomography, X-ray computed  Material decomposition technique  Renal cyst  Pseudoenhancement
基金项目:吉林省科技发展计划项目(20160101074JC)。
作者单位E-mail
刘宏 吉林大学中日联谊医院放射科, 吉林 长春 130031  
万泰虎 吉林大学中日联谊医院放射科, 吉林 长春 130031  
王永亮 吉林大学中日联谊医院放射科, 吉林 长春 130031  
赵卓 吉林大学中日联谊医院手术室, 吉林 长春 130031  
柳林 吉林大学中日联谊医院放射科, 吉林 长春 130031  
乔月 吉林大学中日联谊医院放射科, 吉林 长春 130031  
张孟超 吉林大学中日联谊医院放射科, 吉林 长春 130031 mengchao56@hotmail.com 
摘要点击次数: 2944
全文下载次数: 1275
中文摘要:
      目的 评估能谱CT物质分离技术克服肾囊肿伪强化的价值。方法 80例肾囊肿患者接受能谱GSI模式平扫及双期增强扫描,共入组75个囊肿,按囊肿大小分为3组,A组(n=25)肾囊肿直径0.5~<1.5 cm、B组(n=25)直径1.5~<2.5 cm、C组(n=25)直径2.5~<3.5 cm。采用GSIViewer分析软件生成碘、水基图,记录单能谱70 keV图像上各期肾囊肿的CT值及碘水基值,比较各组肾囊肿平扫和增强双期CT值、碘水基值、CT值和碘水基值差值的差异。结果 3组间实质期与平扫CT值差值总体差异有统计学意义(F=204.128,P<0.001),且两两比较差异均有统计学意义(P均<0.001)。增强后A、B组实质期CT值增加均超过10HU,出现了伪强化现象,C组CT值增加小于10HU,无伪强化现象。各组肾囊肿平扫、皮质期及实质期间碘基值差异有统计学意义(P均<0.001)。3组平扫和增强双期之间碘基值的差值均在10(100 μg/cm3)内,且差值平均值A组> B组> C组(P均<0.05)。A组肾囊肿皮质期及实质期水基值低于平扫(P均<0.05),但差值均在10 mg/cm3以内。结论 能谱CT物质分离技术测定肾囊肿各期碘、水基值时同样会发生碘水基值的"漂移",肾囊肿越小其发生"漂移"的程度越大,碘基值发生漂移的程度受肾囊肿大小的影响较水基值更明显。
英文摘要:
      Objective To evaluate the value of material decomposition imaging of spectrum CT in overcoming renal cyst pseudoenhancement. Methods Totally 80 patients with renal cysts (total 75 cysts) who underwent CT imaging with GSI mode were collected. The renal cysts were divided into 3 groups according to diameters, group A (diameters 0.5-<1.5 cm, n=25), B (1.5-<2.5 cm, n=25) and C (2.5-<3.5 cm, n=25) respectively. The iodine-water density imaging was reconstructed by using the GSI Viewer analysis software. The CT value and iodine-water concentration of the cysts were recorded. The difference of CT value, iodine-water concentration in unenhanced and enhanced dual phases in each group was compared. Results The difference of CT value between plain scan and parenchyma phase among the 3 groups had statistically significant difference (F=204.128, P<0.001), and the differences comparing any two were statistically significant (all P<0.05). The postcontrast attenuation increased more than 10 HU in group A and B, indicating renal cyst pseudoenhancement, and less than 10 HU in group C, which had no pseudoenhancement. There were statistical difference in iodine concentration of the cysts of the 3 groups in unenhance, cortical and parenchyma phases (all P<0.001), but the difference value in unenhance, cortical and parenchyma phases were less than 10 (100 μg/cm3), and the difference value of the 3 group was group A> group B> group C (all P<0.05). The water concentration of the cysts in group A descend in renal cortical and parenchyma phase with statistical difference (P<0.001), but the difference value was less than 10 mg/cm3. Conclusion The measurements of iodine-water concentration appear to drift as well, the smaller the greater. The degree of the iodine concentration shifting is more obvious than water concentration.