张基永,张武哲,陆佳扬,吴丽丽.千伏级锥形束CT计算宫颈癌放疗剂量[J].中国医学影像技术,2015,31(10):1592~1595
千伏级锥形束CT计算宫颈癌放疗剂量
Dose calculations on kilovoltage cone beam computed tomography in radiation therapy for cervical cancer patients
投稿时间:2015-02-09  修订日期:2015-08-13
DOI:10.13929/j.1003-3289.2015.10.033
中文关键词:  体层摄影术,X线计算机  放射治疗剂量  宫颈癌
英文关键词:Tomography, X-ray computed  Radiotherapy dosage  Cervical cancer
基金项目:
作者单位E-mail
张基永 汕头大学医学院附属肿瘤医院放疗科, 广东 汕头 515031  
张武哲 汕头大学医学院附属肿瘤医院放疗科, 广东 汕头 515031  
陆佳扬 汕头大学医学院附属肿瘤医院放疗科, 广东 汕头 515031  
吴丽丽 汕头大学医学院附属肿瘤医院放疗科, 广东 汕头 515031 liliwustu@gmail.com 
摘要点击次数: 1453
全文下载次数: 1090
中文摘要:
      目的 探讨宫颈癌放疗中采用加速器机载千伏级锥形束CT(kV-CBCT)影像进行剂量计算的准确性和可行性。方法 分别于CT模拟机和kV-CBCT上扫描CIRS-062电子密度模体,获得电子密度模体影像及计划CT(pCT)和kV-CBCT的CT值-相对电子密度曲线,采用治疗计划系统计算剂量。选取9例宫颈癌患者的调强治疗计划,并将计划移植至模体和患者的pCT和kV-CBCT影像进行剂量计算,比较剂量差异及剂量分布情况。结果 kV-CBCT的CT值经多次重复测量最大变化≤3%;kV-CBCT的CT值-相对电子密度曲线与pCT相比,低密度组织的差异较小,高密度者差异相对较大。在对模体的剂量计算中,pCT和kV-CBCT的剂量差异为0.7%~2.5%,平均(1.23±0.55)%;而在对宫颈癌患者的剂量计算中,剂量差异为0.5%~1.8%,平均(0.84±0.44)%;剂量分布的一致性均较好。结论 宫颈癌放疗中采用kV-CBCT影像进行剂量计算是可行的。
英文摘要:
      Objective To explore the accuracy and feasibility of dose calculations using kilovoltage cone beam computed tomography (kV-CBCT) in cervical cancer radiotherapy. Methods The CIRS-062 calibration phantom underwent scanning by planning CT (pCT) and kV-CBCT, and then the CT value-electron density curves were obtained. The CT value-electron density curves were inputted to the treatment planning system. Nine intensity-modulated radiation therapy plans of cervical cancer were chosen, and the plans were transferred to the pCT and kV-CBCT data sets for dose calculations. Phantom and patient studies were carried out. The dose differences and dose distributions were compared between kV-CBCT plan and pCT plan. Results The CT value of kV-CBCT was measured several times, and the maximum change was ≤3%. To compare with pCT, the CT value of kV-CBCT had only a little discrepancy for soft tissue, meanwhile it was bigger for hard tissue. For phantom study, the mean dose difference between pCT and kV-CBCT was 1.23%±0.55% (range 0.7%—2.5%). For dose calculation in patient kV-CBCT images, the mean dose difference between pCT and kV-CBCT was 0.84%±0.44% (range 0.5%—1.8%), and the dose distributions agreed very well. Conclusion The kV-CBCT for dose calculation is feasible in cervical cancer radiotherapy.
查看全文  查看/发表评论  下载PDF阅读器