王予生,过哲,赵海竹,刘庆华,白荣杰,程晓光.不同管电压对定量CT骨密度测量结果的影响[J].中国医学影像技术,2015,31(10):1475~1478
不同管电压对定量CT骨密度测量结果的影响
Effect of different tube voltage on result of bone mineral density measured by quantitative computed tomography
投稿时间:2015-03-09  修订日期:2015-06-10
DOI:10.13929/j.1003-3289.2015.10.007
中文关键词:  管电压  骨密度  体层摄影术,X线计算机
英文关键词:Tube voltage  Bone mineral density  Tomography, X-ray computed
基金项目:国家自然科学基金(81071131)、首都卫生发展科研专项自主创新项目(2014-2-1122)。
作者单位E-mail
王予生 北京积水潭医院放射科, 北京 100035  
过哲 北京积水潭医院放射科, 北京 100035  
赵海竹 北京积水潭医院放射科, 北京 100035  
刘庆华 北京积水潭医院放射科, 北京 100035  
白荣杰 北京积水潭医院放射科, 北京 100035  
程晓光 北京积水潭医院放射科, 北京 100035 xiao65@263.net 
摘要点击次数: 1880
全文下载次数: 1173
中文摘要:
      目的 探讨不同管电压对定量CT(QCT)骨密度(BMD)测量结果的影响。方法 采用Toshiba Aquilion 64排CT机,对欧洲标准脊柱体模和QCT校准体模行CT扫描,根据不同管电压分为80 kV组、100 kV组、120 kV组、135 kV组。每组除电压不同外,其余扫描参数相同,重建标准体模影像,并进行BMD测量。结果 4组标准体模平均BMD值差异有统计学意义(F=3944.32,P<0.05);80 kV组、100 kV组、135 kV组平均BMD值与标准体模BMD真值(116. 67 mg/cm3)差异有统计学意义(P均<0.05)。80 kV组、100 kV组、120 kV组、135 kV组BMD测量值减真值的绝对值分别为(57.22±1.80)mg/cm3、(22.82±0.65)mg/cm3、(0.08±0.27)mg/cm3、(15.94 ±2.52)mg/cm3。管电压与平均BMD呈负相关(r=-0.993,P<0.05)。结论 不同管电压对QCT BMD的测量结果有影响。采用QCT测量BMD时,建议将扫描协议中管电压固定为120 kV。
英文摘要:
      Objective To assess the effect of different tube voltage on the quantitative computed tomography (QCT) bone mineral density (BMD) measurement. Methods European Spine Phantom (ESP) and QCT calibration phantom were scanned with a 64-row Toshiba CT scanner under 4 different tube voltages 80 kV, 100 kV, 120 kV, and 135 kV group respectively. In addition to the voltage, the other scanning parameters were the same in different groups. Standard phantom images were reconstructed, and the BMD was measured. Results There were statistically significant differences among the 4 groups of standard phantom mean BMD values (F=3944.32, P<0.05). The BMD of 80 kV group, 100 kV group, 135 kV group were statistically significant compared with that of the standard phantom (BMD=116. 67 mg/cm3, all P<0.05). The absolute value of BMD measured in 80 kV, 100 kV, 120 kV, 135 kV minus the true value were (57.22±1.80) mg/cm3, (22.82±0.65) mg/cm3, (0.08±0.27) mg/cm3, (15.94±2.52) mg/cm3. Tube voltage was negatively correlated with BMD (r=-0.993, P<0.05). Conclusion Different tube voltage can affect BMD measurement of QCT. Tube voltage in QCT scan protocol should be fixed at 120 kV to ensure accurate measurement of QCT.
查看全文  查看/发表评论  下载PDF阅读器