李昂,罗松,沈君姝,郭辉,潘永泉,曹新志.模拟婴幼儿头颅70 kV CT扫描的图像质量和辐射剂量[J].中国医学影像技术,2014,30(9):1424~1427
模拟婴幼儿头颅70 kV CT扫描的图像质量和辐射剂量
Image quality and radiation dose of simulating infant head CT scanning based on 70 kV
投稿时间:2014-04-03  修订日期:2014-07-18
DOI:
中文关键词:  体层摄影术,X线计算机  婴儿  儿童  图像质量  辐射剂量
英文关键词:Tomography, X-ray computed  Infant  Child  Image quality  Radiation dose
基金项目:
作者单位E-mail
李昂 南京军区南京总医院医学影像科, 江苏 南京 210002  
罗松 南京军区南京总医院医学影像科, 江苏 南京 210002  
沈君姝 南京军区南京总医院放疗科, 江苏 南京 210002  
郭辉 南京军区南京总医院医学影像科, 江苏 南京 210002  
潘永泉 南京军区南京总医院医学影像科, 江苏 南京 210002  
曹新志 南京军区南京总医院医学影像科, 江苏 南京 210002
南京航空航天大学生物医学工程系, 江苏 南京 210016 
to_cxz@163.com 
摘要点击次数: 2444
全文下载次数: 1151
中文摘要:
      目的 探讨模拟婴幼儿头颅70 kV CT扫描条件下的图像质量和辐射剂量。方法 选用CPT500模体,分别于70 kV及120 kV、150~500 mAs条件下进行扫描,测量其低对比度模块中对比度为1.0%、直径为15 mm的圆柱体的CNR和CTDIvol值,分析两者的相关性。由2名经验丰富的放射科医师对图像的低对比可探测能力(LCD)进行评分。对70 kV、150~500 mAs测得的每组图像的CNR和LCD评分与婴幼儿头颅常规扫描条件(120 kV、200 mAs)图像的差异行统计学分析。结果 当管电压为70 kV及120 kV时,图像CNR与CTDIvol呈线性正相关性(r=0.96、0.94,P<0.001);当管电流设置为150~400 mAs时,70 kV与常规扫描条件图像的CNR及LCD评分的差异均有统计学意义(P均<0.05),但当管电流设置为450~500 mAs时,差异均无统计学意义(P均>0.05)。结论 70 kV、450~500 mAs扫描条件可得到与传统的120 kV、200 mAs CT扫描相似的图像质量,前者辐射剂量明显降低,适用于婴幼儿头颅CT扫描。
英文摘要:
      Objective To explore the image quality and radiation dose of simulating infant head CT examination based on 70 kV. Methods The CPT500 phantom was scanned in 70 kV and 120 kV, 150-500 mAs, respectively, and the CNR and CT dose index volume (CTDIvol) value of low-contrast-module cylinder (contrast ratio 1.0%, diameter 15 mm) were measured. Two experienced radiologists evaluated the low contrast detection capabilities of image (LCD). And the correlation between CNR and CTDIvol was analyzed. The differences of CNR and LCD between two conditions (70 kV, 150-500 mAs and 120 kV, 200 mAs) was analyzed. Results There were positive correlation between CNR and CTDIvol in condition of 70 kV and 120 kV (r=0.96, 0.94, P<0.001). The differences of CNR and LCD was statistically significant between the images obtained with 70 kV, 150-400 mAs and 120 kV, 200 mAs (all P<0.05). However, in the condition of 70 kV, 450-500 mAs, there was no statistical difference was found (both P>0.05). Conclusion The equal image quality can be obtained in the CT scan condition of 70 kV, 450-500 mAs and 120 kV, 200 mAs, and the former with lower radiation dose is suitable for infant head CT examination.