涂宁,文之,朱凌,刘博,冯晓波,汪长银,吴光耀.双能CT评估骨密度[J].中国医学影像技术,2014,30(9):1391~1394 |
双能CT评估骨密度 |
Dual-energy CT in assessment of bone mineral density |
投稿时间:2014-03-05 修订日期:2014-06-25 |
DOI: |
中文关键词: 放射摄影术 体层摄影术,X线计算机 骨密度 |
英文关键词:Radiography Tomography, X-ray computed Bone density |
基金项目: |
|
摘要点击次数: 2532 |
全文下载次数: 1020 |
中文摘要: |
目的 评价利用双能CT(DECT)测量容积CT值评估骨密度(BMD)的可行性。方法 对11名志愿者采用双能X线吸收法(DXA)测量腰椎(L1~L4)及股骨(股骨颈)BMD值;并对所有研究对象进行DECT检查,获得钙值图,测量容积CT值,分析容积CT值与BMD值的相关性。结果 腰椎容积CT值与BMD值呈显著正相关(r=0.82,P<0.001;相关函数为Y=0.004X-0.879,其中Y为BMD值,X为容积CT值);股骨容积CT值与DXA所测BMD值无明显相关性(P>0.05)。结论 DECT所测腰椎容积CT值与DXA所测BMD值密切相关,可定量反映腰椎BMD变化。 |
英文摘要: |
Objective To evaluate the feasibility of dual-energy CT (DECT) based measurement of volume CT value in assessing bone mineral density (BMD). Methods Totally 11 volunteers were enrolled and underwent dual X-ray absorptiometry (DXA) to obtain BMD and DECT to obtain calcium value map for measurement of volume CT value in lumbar spines (L1-L4) and femoral necks. The correlation of volume CT value and BMD was analyzed. Results The volume CT value was positively correlated with BMD (r=0.82, P<0.001) in lumbar spines, with correlation function of Y=0.004X-0.8791 (Y: DXA BMD value, X: volume value). The volume CT value was not significantly correlated with BMD (P>0.05) in femoral necks. Conclusion Volume CT values measured by DECT is closely correlated with BMD value measured by DXA in lumbar spines, which can indicate quantitatively BMD. |