贾姝妮,陈晓燕,薛继平,康春松,张燕霞,李帅.超声剪切波弹性成像评价成人正常甲状腺的重复性[J].中国医学影像技术,2014,30(6):841~845
超声剪切波弹性成像评价成人正常甲状腺的重复性
Reproducibility of shear wave elastography in evaluation on thyroid of healthy adult volunteers
投稿时间:2013-12-05  修订日期:2014-05-02
DOI:
中文关键词:  超声检查  甲状腺  剪切波  弹性成像技术  可重复性,结果
英文关键词:Ultrasonography  Thyroid gland  Shear wave  Elasticity imaging techniques  Reproducibility of results
基金项目:山西省国际科技合作项目(2013081061)。
作者单位E-mail
贾姝妮 山西医学科学院 山西大医院超声科, 山西 太原 030032  
陈晓燕 山西医学科学院 山西大医院超声科, 山西 太原 030032  
薛继平 山西医学科学院 山西大医院超声科, 山西 太原 030032  
康春松 山西医学科学院 山西大医院超声科, 山西 太原 030032 kangchunsong2005@sina.com 
张燕霞 山西医学科学院 山西大医院超声科, 山西 太原 030032  
李帅 山西医学科学院 山西大医院超声科, 山西 太原 030032  
摘要点击次数: 2987
全文下载次数: 1329
中文摘要:
      目的 评价剪切波弹性成像(SWE)技术检测正常甲状腺组织弹性参数的重复性。方法 选取健康成年志愿者28名,应用SWE技术,由同一检查者于不同时间及不同检查者分别行甲状腺弹性成像,应用Q-Box获取甲状腺双侧叶中部的弹性参数(Emean),评价组内及组间的重复性。结果 ①检查者内同一天两测值Emean差异无统计学意义(P>0.05),甲状腺左、右叶的组内相关系数(ICC)值分别为 0.81、0.94,Pearson相关系数分别为r=0.77、0.88(P均<0.001),提示组内重复性很好;②检查者内不同天两测值Emean 差异无统计学意义(P>0.05),组内ICC值分别为0.59、0.86,Pearson相关系数分别r=0.69、0.72(P均<0.001),组内重复性好;③检查者间两测值Emean差异无统计学意义(P>0.05),左、右叶ICC值分别为 0.64、0.58,Pearson相关系数分别为r=0.45(P=0.014)、r=0.57(P=0.002),组间重复性好。Bland-Altman图示组内及组间SWE测量值具有良好一致性。结论 SWE可定量获取甲状腺的杨氏模量值,重复性好。
英文摘要:
      Objective To assess the reliability of shear wave elastography (SWE) in evaluation on thyroid of healthy volunteers. Methods Totally 28 healthy adult volunteers underwent SWE of thyroid by different inspectors or the same inspector in different time points. Bilateral thyroid lobe central elastic parameters (Emean) were examined with Q-Box, and the intra-observer and inter-observer reproducibility were assessed. Results ①Paired t test showed no significant difference between the two measured values made by same inspectors on the same day (P>0.05), the intra-class correlation coefficient (ICC) values of left lobe and right lobe was 0.81 and 0.94, respectively. Pearson correlation coefficients were 0.77, 0.88 (all P<0.001), indicating very good reproducibility. ②Paired t test showed no significant difference between the two measured values made by the same inspectors on different days (P>0.05), the intra-class ICC values of left lobe and right lobe was 0.59 and 0.86, respectively. Pearson correlation coefficients were 0.69, 0.72 (all P<0.001), indicating good reproducibility.③Paired t test showed no significant difference between the two measured values made by different inspectors (P>0.05), the intra-class ICC values of left lobe and the right lobe was 0.64 and 0.58, respectively. Pearson coefficients, rleft=0.45 (Pleft=0.014), rright=0.57 (Pright=0.002), indicated good reproducibility. Inter-observer and intra-observer Bland-Altman plots also showed good reproducibility. Conclusion SWE has good repeatability to obtain Young's modulus value of normal thyroid.