张健,杨明,莫绪明,靳激扬,刘斌,黎玲玲,滕皋军.不同管电压64层螺旋CT对儿童CTA图像质量和辐射剂量的影响[J].中国医学影像技术,2012,28(6):1213~1217
不同管电压64层螺旋CT对儿童CTA图像质量和辐射剂量的影响
Impact of different tube voltage protocols on image quality and radiation dosage for pediatric 64-slice cardiovascular CT angiography
投稿时间:2011-11-16  修订日期:2012-02-06
DOI:
中文关键词:  儿童  心脏缺损,先天性  体层摄影术,X线计算机  血管造影术  图像质量  辐射剂量
英文关键词:Child  Heart defects, congenital  Tomography, X-ray computed  Angiography  Image quality  Radiation dosage
基金项目:
作者单位E-mail
张健 东南大学附属中大医院放射科,江苏 南京 210009  
杨明 东南大学附属中大医院放射科,江苏 南京 210009 yangming19710217@163.com 
莫绪明 南京医科大学附属南京儿童医院心胸外科,江苏 南京 210008  
靳激扬 东南大学附属中大医院放射科,江苏 南京 210009  
刘斌 东南大学附属中大医院放射科,江苏 南京 210009  
黎玲玲 东南大学附属中大医院放射科,江苏 南京 210009  
滕皋军 东南大学附属中大医院放射科,江苏 南京 210009  
摘要点击次数: 2239
全文下载次数: 1126
中文摘要:
       目的 用64层螺旋CT观察不同管电压(80 kVp及100 kVp)对儿童CTA图像质量(IQ)和辐射剂量(ED)的影响。方法 回顾性分析52例先天性心脏病患儿心脏CTA检查资料,根据扫描管电压分为80 kVp组(n=30)及100 kVp组(n=22),均采用自动管电流调制技术。采用bolus tracking技术自动注射对比剂,记录增强后心腔和大血管的密度,测量背景噪声,计算SNR、对比噪声比(CNR)及有效剂量(ED)。由2名医师采用单盲法评估图像质量。结果 80 kVp和100 kVp两组心腔大血管平均密度、SNR及CNR差异均无统计学意义(P均>0.05);IQ评分分别为4.17±0.70、4.23±0.53(P=0.834);80 kVp组平均ED为(0.458±0.085)mSv,低于100 kVp组的(1.074±0.290)mSv(P<0.001)。结论 儿童心脏CTA检查中,与100 kVp相比,80 kVp管电压可明显降低ED而不影响IQ。
英文摘要:
      Objective To evaluate the impact of 80 and 100 kVp tube voltage on image quality (IQ) and radiation dosage in pediatric cardiovascular CT angiography (CTA) with 64-slice CT. Methods CTA was performed in 52 children with suspected congenital heart disease. Children were randomly assigned to 80 kVp (n=30) or 100 kVp (n=22) group. Other scanning parameters were kept constant. Contrast medium was injected automatically with bolus tracking. Enhancement of heart chamber and great vessels and SNR, contrast-to-noise ratio (CNR) and effective dose (ED) were calculated. Subjective vessel contrast was assessed by two radiologists in consensus. Results There was no statistical difference of mean vessel attenuation, SNR and CNR between the two groups (all P>0.05), nor of subjective IQ (4.17±0.70 vs 4.23±0.53, P=0.834). Mean ED for 80 kVp group ([0.458±0.085]mSv) was significantly lower than that for 100 kVp group ([1.074±0.290]mSv, P<0.001). Conclusion Reduction of tube voltage from 100 to 80 kVp can result in significant decrease of ED in pediatric cardiovascular CTA without deteriorating image quality.