姚志萍,杨昆,汪建涛,任秋实.磁共振成像射频线圈的设计及仿真[J].中国医学影像技术,2006,22(9):1421~1423 |
磁共振成像射频线圈的设计及仿真 |
Design and simulation of the radio-frequency coils of magnetic resonance imaging |
投稿时间:2006-04-27 修订日期:2006-08-16 |
DOI: |
中文关键词: 磁共振成像 射频 笼式线圈 仿真 |
英文关键词:Magnetic resonance imaging Radio-frequency Birdcage coil Simulation |
基金项目:上海市科委重点资助项目 (03DZ19716)。 |
|
摘要点击次数: 3031 |
全文下载次数: 3673 |
中文摘要: |
目的 探讨0.3T永磁MRI系统中笼式线圈的计算机模拟和优化。方法 运用毕奥-萨伐尔定律,通过计算分析,设计了针对3H,129Xe共振频率的8列高通笼式线圈,利用IDL软件进行相关计算和仿真。结果 仿真结果证明了在低场条件下所设计的8列笼式高通线圈产生的射频磁场的均匀性和对称性。当列线数目减少为6列及4列时,笼式线圈产生的磁场均匀性和对称性大大降低。结论 仿真精确度较好,误差较低。线圈列数的降低对B1场均匀性造成破坏。 |
英文摘要: |
Objective To investigate the simulation and optimization of birdcage coil in the 0.3T Permanent Magnet MRI system. Methods The Biot-Savart's law was applied, a high-pass 8-leg birdcage coil, whose resonant frequency was consistent with 3H and 129Xe, was designed through some related calculations and simulations. IDL was used to calculate and simulate our scheme. Results The results showed the uniformity and symmetry of magnetic filed produced by the high-pass 8-leg birdcage coil in a low field. When the legs were reduced to 6 and 4, the uniformity and symmetry fell largely. Conclusion The precision of simulation was good. The uniformity and symmetry of B1 filed would be damaged when the number of the coil legs decreased. |